COATRON
M4
This new manual Coagulation
analyzer measure all clotting based Coagulation tests
with a quarter of the regulary volumes.
Example: 25µl sample + 50 µl
reagent
The Coefficient of Variation (CV) is <2-5%
depending on test and reagent used.
A new double-micro cuvette allows an effective and
quick work, even for more samples.
A unique automatic amplification is especially designed
for all new sensitive Thromboplastin reagents and
equalize high differences of turbidity from reagents
and samples. All specific test parameters are stored
in the system.
The Coatron M4 fits perfect into medium size laboratories.
For the usage in Emergency laboratories it is fairly
suitable.
Tests
PT, APTT, TT, Fibrinogen (derived and Clauss),
Venom Time, AT-III, Factor II, V, VII, VIII, IX, X,
XI, XII, Protein C, Protein S (Coag. method) STAT
(2xPT and 2x APTT)
Selectable dimensions: s, %, Ratio,
INR, Units, (mg/dl)
MORE
INFORMATION
Technical Description
The Coatron M4 is a micro-computer controled Coagulation
instrument with two channels and a LC-Display. The
tests will be selected through keys on a foil keyboard.
The dialog language is user friendly and selectable
in English and German. Other languages can be provided
optionally.
The Main program offers the following menus:
ANALYSIS
SETUP-TEST
SETUP-SYSTEM
SERVICE
The measurement can be started manually by pressing
the OPTIC-key in the same time when reagent is added
or with a connected Autopipette. The results are displayed
in the measured dimension and can be converted into
other desired dimensions by pressing the UNIT-key.
A connected printer will automatically print out the
results in all dimensions with Optic channel and test
name. A patient identification No. can be optionally
selected between "NO PAT.ID", "MANUAL
ENTRY" and "AUTO-SERIES" with Start
No.
Using the TECO Softwaretools the measuring curve
data can be transmitted to a Persoanl Computer and
loaded by any spreadsheet with graphic function. Software
updayes are realized with an easy download from a
computer to the instrument.
Technical Data
Device:
SMD Board
V-25 Microcomputer
18-bit Analog-Digital Converter
1 MB Flash-EPROM
1 MB E2 PROM
Power Supply (external):
Input voltage: 94 to 243 Vac
Input frequency: 50 to 60 Hz
Power output (max) 40 Va
CSA, UL, TÜV-GS, CE-Mark
Optic block:
2-channel Optics
LED with 450 nm
Keyboard: Foil keyboard, 3x8 matrix
with Test-, functions- and Numerical keys.
Display:
4 lines x 20 Characters
Liquid Crystal Display
Incubation block:
24 prewarming positions
4 measuring positions
1 Reagent position, mixed with different adapters
1 Reagent position Ø 16 mm
1 Reagent position Ø 22,5 mm
Temperature: 37,0 °C ± 0,2 °C
Autopipette: (optional)
4-stage Autopipette with electronical start
Volume: 25/50/100/200 µl
Thermal Printer: (optional)
external, serial printer with Print-out of all results
Print-out of all results
Print-out of Test parameters
Print-out of System Test parameters
Print-out of Service Test parameters
Standard Package:
| 1 Pc |
Coatron M2 |
| 1 Pc |
Power Supply |
| 1 Pcs |
Mains cable |
| 25 Pcs |
Double cuvettes |
| 5 Pcs |
Reagent container |
| 1 Pc |
Magnetic stirrer |
| 1 Pc |
Reagent adaptor |
| 1 Pc |
printer cable |
| 1 Pc |
Pipette fastener |
| 1 Pcs |
operation Manual |
| 1 Pcs |
Warranty Card |
| 1 Pcs |
Installation report |
| 1 Pc |
Service report |
Consumables and Accessory:
| |
|
| 19 000 02 |
Double cuvette |
| 19 029 00 |
4-stage Autopipette |
| 19 030 01 |
Reagent adaptor Ø22,5 |
| 19 030 02 |
Reagent adaptor Ø22,8 |
| 19 030 03 |
Reagent adaptor Ø24,2 |
| 19 030 04 |
Reagent adaptor Ø27,8 |
| 80 050 06 |
Stirring magnets |
| 80 525 00 |
Reagent container |
| 80 530 00 |
Reagent tubes |
| 80 535 00 |
Thermal printer |
| 80 920 07 |
Pipette tips |
| 80 920 12 |
Thermal paper |
| 80 920 14 |
Printer cable |
Reagents
TEClot reagents (PT, APTT, TT, FIB, VT, LA,
PC, PS)
TEClot Normal and Abnormal
TEControl N, A1, A2, LA+, PC, PS etc.
Deficient Plasmas
|